Instrument type Bench-top Cardiac Biomarker Analyzer
Throughput Up to 6 samples or assays per run
Measuring time  <17 minutes for up to six simultaneous samples
Sampling material Whole blood, plasma
Measuring principle Chemiluninescence enzyme immunoassay technology (CLEIA) combined with proprietary Magtration™ technology
Reaction temperature 37.5º C
Sample volume 100µL
Wavelength 300 – 650 nm
Data storage Patient data: 1000
QC data 1800
CAL data 300
Data transfer ASTM standard
Dimensions 13.5 in. W x 22.4 in. D x 18.7 in. H
Weight 62 lbs
Electrical requirements 100 – 240 Vac
Monitor/keyboard LCD Touchscreen
Printer Integrated
PC Integrated
Interface RS-232C
Calibration Factory calibration, 2-point calibration every 4 weeks
24-h operation (standby) 24-h operation (standby)